An Affordable Calibrator that Combines Three Functions in One
The Calog-Temp, Multi-Function Temperature Calibrator is a locally designed and manufactured, high-precision, hand-held calibrator for the calibration and troubleshooting of process control instrumentation.
It can simultaneously measure and source ten types of thermocouples and seven resistance temperature detectors (RTD), whilst automatically capturing the calibration results. It measures milliamps whilst sourcing or simulating millivolts, TCs, RTDs or ohms to calibrate transmitters. In addition, it has the functionality necessary to install and maintain all powered and non-powered transducers using the built-in 24V power supply milliamp (mA) loops.
The trend feature is ideal for graphing temperature profiles and PID controller optimization with a programmable timebase.
The Calog-Temp can store up to ………… readings in its logging mode which can then be downloaded via an SD card to a PC via the onboard USB port. The data can then be opened using most of the common spreadsheet formats. The document procedures and results all meet ISO 9001/2000 and CE requirements and they come standard with rechargeable batteries in a kit case with test leads, a desktop charger and a rubber boot.
The Calog-Temp is the economical choice for people wanting to do industrial field or workshop calibration and offers a real alternative to pricey internationally recognized brands of multi-function calibrators. Whether you need to calibrate or troubleshoot RTDs, Thermocouples or mA Loops , the Calog-Temp is the instrument for you.
MEASURING |
ANALOGUE INPUT RANGES |
IMPEDANCE |
ACCURACY |
RESOLUTION |
0 to 24mA |
Input impedance ±17Ω |
0.02% FS |
1µA |
-10 to 100mV |
Input impedance > 1MΩ |
0.01% FS |
1µA |
Thermocouple types: K, J, T B, R, S, E, N, U, L |
Input impedance > 1MΩ |
0.01% FS* |
0.1ºC |
* (Using Thermocouple Extender and CJ Temp set to External) |
0 to 400.00Ω |
|
0.05%FS |
0.01Ω |
0 to 2200.0Ω |
|
0.05%FS |
0.1Ω |
RTD Types: |
Pt50, Pt100, Pt200, Pt500, and Pt1000 |
|
Ni100 and Ni120 |
|
|
0.05%FS |
0.01ºC |
Measuring of Ohms and RTD in 2, 3, or 4-wire configuration |
Continuity with a 100Ω trigger. Visual and audible confirmation |
|
|
|
|
SOURCING |
ANALOGUE INPUT RANGES |
MAX LOAD |
ACCURACY |
RESOLUTION |
0 to 24mA |
Output load max 500Ω |
0.02% FS |
1µA |
-10 to 100mV |
Min. load 100Ω |
0.01% FS |
1µA |
Thermocouple types: K, J, T B, R, S, E, N, U, L |
|
0.01% FS* |
0.1ºC |
* (Using Thermocouple Extender and CJ Temp set to External) |
|
0 to 400.00Ω |
10 - 400Ω |
0.05%FS |
0.01Ω |
0 to 2200.0Ω |
10 - 2k2Ω |
0.05%FS |
0.1Ω |
RTD Types: |
Pt50, Pt100, Pt200, Pt500, and Pt1000 |
|
Ni100 and Ni120 |
|
|
0.05%FS |
0.01ºC |
|
|
|
|
ISOLATION |
|
|
|
The mV/TC/Ohms/RTD output from the CALOG - TEMP is isolated to a miximum rating of 100VDC from mA measuring/source. |
|
|
|
|
INCLUDED ACCESSORIES |
RECOMMENDED ACCESSORIES |
HW447 - Red Safety Test Lead |
RAL-CIG - 12Vdc Charger (Car Lighter) |
HW448 - Black Safety Test Lead |
|
BA042 - Li-ion Battery Pack |
|
HW418 + HW422 - Mains Charger |
|
HW033 - Rubber Boot |
|
HW180-A - Plastic Kit Case |
|